Kippzonen COMBILOG Data Logger Manuel d'utilisateur Page 148

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 173
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 147
13 SPECIFICATIONS
13.1 Power Supply
Voltage range: +10 ... 30 VDC
Power input: depending on scan rate
approx. 240 mW at measuring rate
60 s = 20.00 mA (at 12 V)
approx. 240 mW at measuring rate
10 s = 20.00 mA (at 12 V)
approx. 264 mW at measuring rate
1 s = 22.00 mA (at 12 V)
(all currents measured with
deactivated LC display and
ethernet but with activated
RS232 / RS485 drivers)
13.2 Signal Inputs/Outputs
Analogue inputs: 8, isolated from processor core
no isolation between channels
Digital inputs/outputs: 6 (direction configurable)
13.3 Signal Processing
Scan rate: 0.25, 1, 2, 3, 4, 5, 10, 20, 30 sec.
1, 2, 3, 4, 5, 10, 20, 30, 60 min.
Averaging interval: 1, 2, 3, 4, 5, 10, 15, 20, 30 sec.
1, 2, 3, 4, 5, 10, 15, 20, 30 min.
1, 2, 3, 4, 6, 8, 12h
A/D-conversion: Delta-Sigma, precision app.16 Bit
Accuracy of the real
time clock: <1 s / day
Hardware Manual COMBILOG 1022